Total aflatoxin ELISA kit

ELISA Test
Total aflatoxin ELISA kit
Total aflatoxin ELISA kit

  1. Purpose
    This kit is Aflatoxin Quantitative detection of residues in feed, fish, shrimp,
    tissue, egg, honey, milk, serum and urine sample.
    2. Principle
    This kit uses a direct competitive ELISA method in microplate coated with
    coupled antigen, adding standard or samples, free and pore strips Pre-coated
    conjugated antigen compete against antibody enzyme marker color with TMB substrate
    stop solution added after the color from blue to yellow, with a microplate reader
    detection wavelength at 4 5 0 nm, the sample absorbance value and content is
    inversely proportional to the sample through the standard curve content.
    3. Kit composed
    ( 1)Pre-coated coupled antigen removable microtiter plates:1 block (12 wells×8 strips)
    (2)Standard:6 bottles(1ml/vial),contents were:0 ng/ml, 0.2 ng/ml,0.6 ng/ml, 1.8
    ng/ml,5.4 ng/ml, 16.2 ng/ml.
    Anti- AFT- B1 antibody conjugate: 1 bottle (6ml)
    (4) Color liquid A: 1 bottle (6ml)
    (5) Color liquid B: 1 bottle (6ml)
    (6) Stop Solution: 1 bottle (6ml), 2M sulfuric acid.
    (7) Sample Diluent: 1 bottle ( 10 × , 6ml), used for sample dilution.
    (8) washing liquid concentrate: 1 bottle (20 × , 20ml), for the washer
    (9) A description
    4. The material needs without providing
    (1) Equipment
    1) Wavelength 450nm microplate reader.
    2) Mill.
    3) Graduated cylinder.
    4) Oscillator.
    5) Funnel.
    6) Whatman No 1 filter paper or equivalent.
    7) Pipette.
    GloryScienceCo.,Ltd
    www.glorybios.com
    (2) Reagent
    1) Deionized water or distilled water.
    2) Methanol.
    5. Notes
    — Using the kit, please carefully read the instructions before
    —Do not use expired kits.
    — Kit Before use, the reagents to room temperature (25 ± 2 C), recommend at least
    2 hours back to temperature.
    —Standard contains AFT- B1, use with caution, the operation should wear gloves.
    — Stop solution contains sulfuric acid, used clothing to prevent skin burns and
    corrosion.
    — Different standards, the sample used in the nozzle can not be mixed, otherwise it
    will affect the test results.
    — Different lots of reagent kit not mix; different standard samples are not mixed with
    the tip, otherwise it will affect the experimental results.
    — Diluted samples must be used in the sample kit diluent, otherwise it will affect the
    experimental results
    — Avoid mixing foaming agent. Kit stored at 2 ~ 8 C, not frozen, not dried out of
    the microplate should be kept sealed
    6. Working fluid
    ( 1) AFT- B1standard solution: 0 ng/ml, 0.2 ng/ml,0.6 ng/ml, 1.8 ng/ml,5.4 ng/ml, 16.2
    ng/ml.
    (2) Concentrated washing solution: distilled water ratio of 1:20 ( 1 + 19) dilution
    reserve
    (3) Sample Diluent: distilled water ratio of 1:10 ( 1 +9) dilution reserve
    (4) Color reaction: is standby to prevent direct rays of light
    (5) Reaction termination solution: have spare
    7. Sample handling procedures
    ( Sample during the extraction process, we must strictly follow operating instructions
    should be accurately dilute the extraction process, it will lead to inaccurate results,
    samples should be stored in a cool dark place and kept refrigerated)
    ( 1) 10g crushed samples taken, add 20ml 70% methanol solution
    (2) Strong 3-minute oscillations
    (3) With Whatman No 1 filter paper
    (4)take 100 apms l treatment of samples, join 400 apms l samples diluent
    (5) take 50μl sample dilutions analysis (sample diluted times for 5) .
    8. Elisa analysis steps
    GloryScienceCo.,Ltd
    www.glorybios.com
    (1) Experimental Notes
    1 ) Keep all reagents before the start of the experiment outside the box to fully restore
    to room temperature (25 ± 2 C), about 2 hours. Warmed to room
    temperature (25 ± 2 C) before removing the porous section, the excess pore
    re-sealed immediately in Article 2 ~ 8 C drying saving
    Note: You must ensure the return to full temperature, or the accuracy of detection
    and accuracy.
    2) immediately after use reagents stored back into the 2 ~ 8 C
    3) Do not change the analysis program
    4) Use the exact pipette
    5 ) Operation once started, please do not interrupt any program
    6 ) ELISA results of the reproducibility of a great extent depends on the operating
    program, strictly in accordance with the requirements of operation
    7 ) To avoid cross- contamination, each standard and sample should be added using a
    different kind of tip
    8 ) Do not add to the tip sample contact with the porous surface of the solution
    (2) Analysis of the steps
    1) Pre-numbered, labeled B0, standard and sample location, detecting holes is
    recommended
    2 ) Take the required amount of micropore ( pore removable section) , the excess back
    into the strip re-sealed and immediately save 2 ~ 8 C
    3) Sample dilution ( 10 ×), concentrated washing liquid (20 ×) was diluted into the
    working fluid
    (distilled or deionized water to dilute)
    4) Hole in the B0 adding 50μl 0.0 ng/ml standard solution
    5 ) Add 50μl of the standard holes in the standard solution
    6) Hole in each sample by adding 50μl sample solution
    7) In all holes by adding 50μl of Anti- AFT- B1 antibody conjugate
    8) Gently shake the plate for several seconds
    (3) 37 temperature bath 30min
    ( Warm bath tap from time to time during the reaction plate, holes can reduce the
    error)
    1 ) Get rid of holes in the liquid, wash with a lotion microplate 5 times, the last one
    should beat the absorbent paper to remove the holes in the liquid completely
    (4) Reaction
    1) The washing process is complete, immediately pipette in each pore of color to join
    the 50μl solution A, add 50μl color liquid B; mild shaking plate to mix thoroughly
    2) 37 C 10min warm bath
    3) Each well add 50μl stop solution, mix
    4) The absorbance at 450nm under the test results be read in 5min.
    GloryScienceCo.,Ltd
    www.glorybios.com
    9. Calculation of results
    (1) Quantitative Analysis
    1) Obtained samples of each concentration of standard solution and the mean
    absorbance value ( B) divided by a standard (0 standard) of the absorbance value (B0)
    and multiplied by 10 0 % , that percentage of the absorbance values. B- standard solution
    or sample solution, the average absorbanceB0 - 0 ng/ ml standard solution of the mean
    absorbance values
    2 ) concentration on the value of X axis, Y axis is percentage of absorbance,
    the standard curve graph. Percentile according to the sample absorbance value can
    be obtained from the corresponding point on the curve the abscissa, that is,
    concentration on the values obtained shall be measured against the number of
    medium concentration C (ng/ml)
    3 ) After a pre-dilution of the sample, so the standard curve derived by multiplying the
    sample concentration dilution factor will again.
    (2 ) Semi- quantitative determination
    1 ) The visual semi-quantitative determination: first select an appropriate standard
    solution and
    sample with the operation, according to the sample absorbance with the standard level
    of comparison, to determine the sample concentration is less than or greater than the
    standard value.
    2) Semi-quantitative measurement instrument: first select an appropriate standard
    solution and
    sample with the operation, according to the color depth of the sample compared with
    the standard
    to determine sample concentration value is less than or greater than the standard
    value.
    10 .specific
    Material cross- reactivity.
    Total aflatoxin 100%
    11. Kit Parameters
    —This kit detection limit 0.2ng/ml.
    —B0 best value should be greater than 1.0 absorbance
    —Absorbance plate kit error of less than 8%, less than 15% between the plates.
    — With the manual extraction of tissue samples provided by more than 8 0 % recovery
    rate.

  2. 12. Standard curve mode
    Kit provided by the standard curve range 0.2ng/ml ~ 16.2ng/ml.
    13 Analysis of restriction
    This kit samples test positive should be another way, such as HPLC or GC / MS